Coding <type of calculation>:
ASCII "0": normal calculation of average value
ASCII "1": calculation of average value with wind direction
ASCII "2": calculation of the sum over the averaging interval
ASCII "3": continuous sum
ASCII "4": vectorial average for wind velocity
ASCII "5": vectorial average for wind direction
events (b...b) length = variable
<year> ASCII 2 char
<month> ASCII 2 char
<day> ASCII 2 char
<hour> ASCII 2 char
<minute> ASCII 2 char
<second> ASCII 2 char
<separation mark> ASCII (“;“) 1 char
<average value channel 1> ASCII 8 char
<separation mark>
1
ASCII (“;“) 1 char
<average value channel n>
1
ASCII 8 char
<separation mark>
1
ASCII (“;“) 1 char
1
The response telegram will only contain these fields if
they have been defined.
The transmission of average values is performed by 8 hexa-
decimal characters in the IEE Std 754 Short Real Format (Ex-
ample: 50,3094 -> 42 49 3C D3).
Depending on the programming language being used, different
terms are used, e.g. format "single" with use of Turbo Pascal.
(refer to appendix H)